E sCO E sCOV E sCAiO E sCAiOV Gas Module Single Wide

E-sCO, E-sCAiO, E-sCAiOV - Gas Module - Single Wide

Model Number: E-sCO, E-sCAiO, E-sCAiOV

The GE E-sCO, E-sCOV, E-sCAIO, E-sCAIOV AGM module can read O2, CO2, and N2O. The E-sCOV also has spirometry. The E-sCIAO has Agent Identification and readings capabilities. The E-sCAIOV has both Agent Identification, readings, and spirometry.    

The GE E-sCAIO AGM Gas module is a very small and compact module. All models can identify and read O2, N20, and CO2. The E-sCOV and E-sCAIOV also have Spirometry. The E-sCAIO and E-sCAIOV can read and identify Agent Gas. The module is single wide making it very easy to implement in many situations. It works with GE module racks and internal module racks on GE Monitors with compatible software. The OEM requires a normal PM and checkout each year. Every four years requires replacement of a CO2 Scrubber which is sold separately from the standard PM kit. Common issues we see are bench or pump failures. Common errors displayed on screen are “SENSOR INOP” and “Gas Module Error.”

Troubleshooting help:

Basic Operation Information Flow rate 120ml/min +/- 20, Calibration Gas: 755571-HEL 5% CO2, 54.5% O2, 36.0% N2O, 2.0% DESFLURANE,BAL N2

Helpful Files:

  • Sensor INOP
  • Occlusion
  • Fails Leak test
  • Intermittent resets
  • Pm kits
  • Check D-fend
  • Calibration

Everything that comes out of our shop has been cleaned, serviced, and is ready for use.

 

Give us a call:

1-888-644-3274